## Limit of Detection and Quantitation

LOD is the lowest analyte concentration that can be reliably detected (S/N ≈ 3). LOQ is the lowest concentration that can be quantified with acceptable precision (S/N ≈ 10).

### Formulas (ICH Method)

**LOD = 3.3 × sigma / S** (often simplified to 3 × sigma / S)

**LOQ = 10 × sigma / S**

where sigma is the standard deviation of the blank response and S is the slope of the calibration curve. This is the most common approach in pharmaceutical and environmental analysis.

Exemple Résolu

Blank standard deviation = 0.005, calibration slope = 2500 signal/M.

  1. LOD = 3 × 0.005 / 2500 = 0.015 / 2500 = 6.0 × 10⁻⁶ M = 6.0 uM
  2. LOQ = 10 × 0.005 / 2500 = 0.05 / 2500 = 2.0 × 10⁻⁵ M = 20 uM